Listed Below
 
Request Scruz News
Thickness Gauges
Flaw Detectors
Bolting Products
Transducers
Transducers
Transducers
Transducers
Transducers
Return to Main Page
Group Logo
 
 
 


Adjustable square wave pulser provides the flexibility necessary for both high resolution and penetration requirements.

Selectable viewing options provide the user with additional flexibility during operation: (RF waveform, +/- Rectified waveform, and Large Digits with Scan Bar.

Time based B-Scan feature displays a cross section of the test material. Displays the profile of the opposite surface of the material.

Adjustable resolution settings add to the PVX’s flexibilty.

Ability to use a variety of single element transducers for specific applications: Standard Delay Line (acrylic and graphite tips for metals and thin plastics), Pencil Delay Line (tough access area on thin materials), and Contact transducers (variety of applications).

Hardware AGC gain control for multiple echo and thru-paint measurement.

Multiple calibration options: One Point, Two Point, or selection from a Material List.

16 factory setups and 48 user-defined setups. User defined setups can be edited for custom applications.

PVX is equipped with an alpha-numeric data logger to provide increased versatility for those custom reporting needs.

The high speed scan feature speeds up the inspection process by making 32 measurements per second. Remove transducer from the test material and display the minimum measurement scanned.

Visual and audible alarm with hi and lo limit settings for specific application tolerances.

Auto find features locates the detection point(s) and adjusts the display settings to bring the waveform into view.

PVX comes complete with our Windows® PC software for transferring data to and from a PC.

2 year limited warranty

 
   Specifications:
 

Physical

Weight:
13.5 ounces (with batteries).

Size:
2.5 W x 6.5 H x 1.24 D inches
(63.5 W x 165 H x 31.5 D mm).

Operating Temperature:
-14° to 140°F (-10° to 60°C).

Keyboard:
Membrane switch with twelve tactile keys.

Case:
Extruded aluminum body
with nickel-plated aluminum
end caps (gasket sealed).

Data Output:
Bi-directional RS232 serial port. Windows® PC interface software.

Display:
1/8 in. VGA grayscale display
(240 x 160 pixels). Viewable area 2.4 x 1.8 in. (62 x 45.7mm).
EL backlit (on/off/auto).

Ultrasonic Specifications

Measurement Modes:
Pulse-Echo (Precision General purpose).

Interface-Echo (Precision - Thick materials).

Echo-Echo (Precision - Thin materials & thru-paint).

Pulser:
Square wave pulser with adjustable
pulse width (spike, thin, wide).

Receiver:
Manual or AGC gain control with
40dB range, depending on mode
selected.

Timing:
40 MHz with ultra low power
8 bit digitizer.














Power Source

Three 1.5V alkaline or 1.2V NiCad AA cells.

Typically operates for 150 hours on
alkaline and 100 hours on NiCad
(charger not included.)

Auto power off if idle 5 min.

Battery status meter.

Measuring

Range:

Interface-Echo Mode: Steel 0.05 1.0 in (1.27mm - 25.4mm). Plastics from .005in. (127mm).

Echo-Echo Mode: Steel .006in -.500in (.15mm - 12.7mm).

Pulse-Echo Contact: Steel .040in - 10.0in (1mm - 254mm). Plastics from .010in (.254mm).

Echo-Echo Contact: Steel thru-paint .100in - 3.0in (2.54mm - 76.2mm).

Resolution: (selectable) +/- .001in (0.01 mm), or +/- .0001in - (.001mm).

Velocity Range:
.0492 to .3936 in./ms 1250 to 9999 meters/sec

Single and Two point calibration option, or selection of basic material types.

Units: English & Metric

Display

Display Views:

A-Scan Rectified +/- (half wave view) RF (full waveform view)

B-Scan Time based cross section view. Display speed of 15 secs per screen.

Large Digits Standard thickness
view. Digit Height: 0.400 in (10mm).

Scan Bar 6 readings per second. Viewable in B-Scan and Large Digit views.

Repeatability Bar Graph Bar graph
indicates stability of reading.










Data Logger
(Internal)

12,000 readings and waveforms (alpha numeric storage).

OBSTRUCT to indicate inaccessible locations.

Memory:
16 megabit non-volatile ram.

Transducer

Transducer Types:

Single Element (1 to 20 MHz).

Locking quick disconnect “ 00” LEMO connector.

Standard 4 foot cable.

Custom transducer and cable lengths available.

Features

Setups:
16 factory and 48 user-defined setups.

Gates:
Single gate in pulse-echo mode.
Single gate with holdoff in interface-echo, echo-echo, and plastics mode. Adjustable threshold.

Multiple Measurement Modes:
Selectable modes for use with a variety of applications.

Alarm Mode:
Set hi and lo tolerances with
audible beeper and visual LEDs.

Fast-Scan Mode:
Takes 32 readings per second and
displays the minimum reading found when the transducer is removed. Display continuously updates while scanning.

Certification

Factory calibration:

traceable to
national standards.

Warranty

2 year limited














   
 
Larger Image
Click on the photo above for a larger image.
 
 
 
 
Powered By:
Dakota Ultrasonics 2007
Live chat by Boldchat
 
ASNT Society
 
NACE Society
 
NDT Solutions
 
SonaTest
 
Balteau X-Ray